Create README.md
Browse files
README.md
ADDED
|
@@ -0,0 +1,59 @@
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 1 |
+
Main protocol
|
| 2 |
+
|
| 3 |
+
Performer: Hamza
|
| 4 |
+
Purpose / Hypothesis: To see pump height effect on signal.
|
| 5 |
+
|
| 6 |
+
Needle size: 22G
|
| 7 |
+
Needle tip: Chiba
|
| 8 |
+
|
| 9 |
+
Used Sensor : M.4.3.2
|
| 10 |
+
Punctured Material: Foam
|
| 11 |
+
Phantom setup: Foam wrapped around tube. Both wet and dry foam used.
|
| 12 |
+
|
| 13 |
+
|
| 14 |
+
|
| 15 |
+

|
| 16 |
+
|
| 17 |
+
Extended:
|
| 18 |
+
|
| 19 |
+
Pump was running at 22 rpm .Pattern box used for stable tube and known patter and distance from tube. 3 different height used. Needle holder and rpi was at another table in order to have no contact with phantom expect needle.
|
| 20 |
+
Table ---> Pump output same level as phantom input
|
| 21 |
+
1st Shelf ---> Pump output was 35 CM higher then phantom input
|
| 22 |
+
2nd Shelf ---> Pump output was 75 CM higher then phantom input
|
| 23 |
+
|
| 24 |
+
|
| 25 |
+
|
| 26 |
+
|
| 27 |
+

|
| 28 |
+
|
| 29 |
+
Table level
|
| 30 |
+
|
| 31 |
+
|
| 32 |
+
|
| 33 |
+

|
| 34 |
+
|
| 35 |
+
1st Shelf Level
|
| 36 |
+
|
| 37 |
+
|
| 38 |
+
|
| 39 |
+

|
| 40 |
+
|
| 41 |
+
2nd Shelf Level
|
| 42 |
+
|
| 43 |
+
|
| 44 |
+
|
| 45 |
+
We used 2 different point on pattern. 5G cell and abovetube. With 5G cell we had 3 different amount of needle shaft inside foam. 1cm , 2cm and 3cm mean how many cm of needle shaft was inside of foam during recording. For abovetube point we used 2 different size which were 1 cm and 2 cm.
|
| 46 |
+
|
| 47 |
+
Filename meaning:
|
| 48 |
+
5G_1CM_Table_Dry_foam_22g_Chiba_M.4.3.2_2025-09-24_12.49.17
|
| 49 |
+
|
| 50 |
+
position-depth of the needle shaft inside foam-level of pump-phantom-needle gauge-needle type-sensor-type_timestamp
|
| 51 |
+
|
| 52 |
+
|
| 53 |
+
Sensor places on channel 0. On channel 1 external microphone for sync sound detecting.
|
| 54 |
+
|
| 55 |
+
Needle tracking protocol
|
| 56 |
+
|
| 57 |
+
There is no tracking because needle was stationary
|
| 58 |
+
|
| 59 |
+
|