Gawron97 commited on
Commit
d0f38fd
·
verified ·
1 Parent(s): e806f42

Update README.md

Browse files
Files changed (1) hide show
  1. README.md +19 -20
README.md CHANGED
@@ -1,56 +1,55 @@
1
  Main protocol
2
 
3
  Performer: Hamza \
4
- Purpose / Hypothesis: To see pump height effect on signal.
5
 
6
  Needle size: 22G
7
  Needle tip: Chiba
8
 
9
- Used Sensor : M.4.3.2
10
- Punctured Material: Foam
11
- Phantom setup: Foam wrapped around tube. Both wet and dry foam used.
12
 
13
 
14
 
15
- ![image](https://cdn-uploads.huggingface.co/production/uploads/6866bd792d33c8c807a05d45/3dr4K9TdqeW31E1Ph_dfn.png)
16
 
17
  Extended:
18
 
19
- Pump was running at 22 rpm .Pattern box used for stable tube and known patter and distance from tube. 3 different height used. Needle holder and rpi was at another table in order to have no contact with phantom expect needle. \
20
- Table ---> Pump output same level as phantom input \
21
- 1st Shelf ---> Pump output was 35 CM higher then phantom input \
22
- 2nd Shelf ---> Pump output was 75 CM higher then phantom input
23
 
24
 
25
 
26
 
27
- ![image](https://cdn-uploads.huggingface.co/production/uploads/6866bd792d33c8c807a05d45/Md7aMm6XnGRYs0iKNkquq.png)
28
 
29
- Table level
30
 
 
31
 
32
 
33
- ![image](https://cdn-uploads.huggingface.co/production/uploads/6866bd792d33c8c807a05d45/-rV1bgGr6WIJN2IfIUwe7.png)
34
 
35
- 1st Shelf Level
36
 
 
37
 
 
38
 
39
- ![image](https://cdn-uploads.huggingface.co/production/uploads/6866bd792d33c8c807a05d45/REblm0ePGxbTVYiz2Ek6_.png)
40
 
41
- 2nd Shelf Level
42
 
43
 
44
-
45
- We used 2 different point on pattern. 5G cell and abovetube. With 5G cell we had 3 different amount of needle shaft inside foam. 1cm , 2cm and 3cm mean how many cm of needle shaft was inside of foam during recording. For abovetube point we used 2 different size which were 1 cm and 2 cm.
 
46
 
47
  Filename meaning: \
48
- 5G_1CM_Table_Dry_foam_22g_Chiba_M.4.3.2_2025-09-24_12.49.17
 
 
49
 
50
- position-depth of the needle shaft inside foam-level of pump-phantom-needle gauge-needle type-sensor-type_timestamp
51
 
 
52
 
53
- Sensor places on channel 0. On channel 1 external microphone for sync sound detecting.
54
 
55
  Needle tracking protocol
56
 
 
1
  Main protocol
2
 
3
  Performer: Hamza \
4
+ Purpose / Hypothesis: To see pump height effect on signal with stationary needles.
5
 
6
  Needle size: 22G
7
  Needle tip: Chiba
8
 
9
+ Used Sensor : M.4.4.0.22G and M.4.4.1.22G \
10
+ Punctured Material: Foam \
11
+ Phantom setup: Foam wrapped around tube. Dry foam used.
12
 
13
 
14
 
15
+ ![b378dd1a-7d64-4041-b208-0cee68f8eafc](https://cdn-uploads.huggingface.co/production/uploads/6866bd792d33c8c807a05d45/psXqFHnBSFDUG9GLgClKx.png)
16
 
17
  Extended:
18
 
19
+ Pump was running at 22 rpm. \
20
+ Pattern box used for stable tube and known pattern and distance from tube. \
21
+ 2 differnt pump height used one was 1.5m and other one was same level as pattern box.
 
22
 
23
 
24
 
25
 
 
26
 
27
+ ![60e634a3-9af3-46c7-a4f4-4944512818dc](https://cdn-uploads.huggingface.co/production/uploads/6866bd792d33c8c807a05d45/Dthzi4RmTObyatX4L4Fzw.png)
28
 
29
+ 1.5M high
30
 
31
 
 
32
 
 
33
 
34
+ ![f2597f2c-281f-4c29-ae07-3095e03752f5](https://cdn-uploads.huggingface.co/production/uploads/6866bd792d33c8c807a05d45/e4nnQX-hJkS8s8hCPB3ED.png)
35
 
36
+ Same level
37
 
 
38
 
 
39
 
40
 
41
+ We used 2 different point on pattern. 5G cell and abovetube.
42
+ With 5G cell we had 3 different amount of needle shaft inside foam. 1cm , 2cm and 3cm mean how many cm of needle shaft was inside of foam during recording.
43
+ For abovetube point we used 2 different size which were 1 cm and 2 cm.
44
 
45
  Filename meaning: \
46
+ 1.5m_elevated_1st__M.4.4.0_2nd_M.4.4.1_dry_foam_22g_chiba_2025-10-21_18.04.09
47
+
48
+ level of pump-sensor-type-phantom-needle gauge-needle type_timestamp
49
 
 
50
 
51
+ M.4.4.0 Sensor placed on channel 0. On channel 1 M.4.4.1 microphone.
52
 
 
53
 
54
  Needle tracking protocol
55