29102 | 20200 | D | 1 | null | A 75-YR-OLD MALE WITH HISTORY OF GUILLAIN BARRE SYNDROME IN THE INTENSIVE CARE UNIT. PT HAD BEEN IN THE HOSP SINCE 9/95. PT FOUND AT APPROX 10:48 AM ON 11/14/95 TO BE DISCONNECTED FROM HIS VENTILATOR FOR AN UNKNOWN PERIOD OF TIME. WHEN FOUND HE HAD CYANOSIS AROUND THE LIPS AND A RATE OF 54. PUPILS WERE APPROX 4 MM AND ... | foitext1996.txt |
29104 | 19148357 | D | 1 | null | PHYSICIAN SHOCKED BY ELECTROSURGICAL PENCIL. | foitext1996.txt |
29143 | 17177348 | D | 1 | null | COMPLAINANT HAD A TOTAL OF 8 PEDICLE SCREWS IMPLANTED TO STABILIZE SPINAL COLUMN AFTER TRAUMATIC INJURY TO HER BACK FOLLOWING A SEVERE CAR ACCIDENT IN 2/94. NOW 2 OF THE SCREWS NEED TO BE REMOVED BECAUSE THEY HAVE BROKEN. | foitext1996.txt |
29146 | 45658 | D | 1 | null | MEDEX, INC. 3637 LACON RD, HILLIARD, OH 43026-1289. THEIR INVESTIGATION REPEATEDLY SHOWED THAT THIS ISSUE RESULTED WHEN THE CLINICIAN INSERTED TEH NEEDLE INTO THE BURRETTE INJECTION PORT AT A SEVERE ANGLE AND SHEARED OFF A SECTION OF THE PLASTIC MATERIAL WITH THE NEEDLE. THIS WAS DISCUSSED WITH THE HOSP STAFF. THE PROD... | foitext1996.txt |
29147 | 17280 | D | 1 | null | THE SYRINGE PUMP WAS BEING USED TO INFUSE MEDICATION IN A 4-YR-OLD PT UNDERGOING A PROCEDURE IN MRI. THE PUMP WAS APPROX 5-6 FEET FROM THE MRI MAGNET WHEN THE FORCE OF THE MAGNETIC FIELD PULLED THE PUMP FROM THE HANDS OF A CRNA. THE PUMP FLEW THROUGH THE AIR AND ATTACHED TO THE MAGNET. THE PUMP DID NOT STRIKE THE CHILD... | foitext1996.txt |
29160 | 17282 | D | 1 | null | PER RPTR, APPARENTLY, THE DONOR BLOOD AND TRANSPORT SOLUTION TESTED POSITIVE FOR PSEUDOMONAS AERUGINOSA, BUT AFTER EXPOSURE TO THE ANTIBIOTIC COCKTAIL, THE COCKTAIL AND CO-PROCESSED TISSUE CULTURED NEGATIVE, SO THE PROCESSOR FELT THAT THE TISSUE WAS STERILE AND RELEASABLE. THERE WERE TWO RECIPIENTS WHO RECEIVED VALVES ... | foitext1996.txt |
29170 | 21388 | D | 1 | null | RODS HOLDING THE LOWER 5 VERTEBRAE ARE HELD IN PLACE BY 4 SCREWS OR BOLTS THAT WERE NOT INTENDED FOR THAT USE. THE SCREWS ARE TOO LARGE AND RUB AGAINST MUSCLES INCAPACITATING THE PT. | foitext1996.txt |
29174 | 17908405 | D | 1 | null | THREE PATIENTS ARE INVOLVED. THE DISCS FAILED TO DIAGNOSE STREP INFECTIONS. THE PATIENTS CONTINUED TO HAVE HIGH FEVERS. TESTS WERE CONDUCTED WITH ANOTHER PRODUCT AND ALL THREE PATIENTS WERE GROUP A STREP POSITIVE. THE RPTR THINKS THE PRODUCT IS DEFECTIVE. | foitext1996.txt |
29177 | 17307 | D | 1 | null | PROVIDERS COMPLAINED OF SPECULUMS BEING SHORT AND THEY WERE DIFFICULT TO LOCK PROPERLY DURING INSERTION. THEY WERE ALSO VERY DIFFICULT TO UNLOCK AFTER A PAP PROCEDURE, INCREASING PT PAIN. | foitext1996.txt |
29181 | 17308 | D | 1 | null | DURING DIAGNOSTIC LAPAROSCOPY THE INSULATING COATING OF THE BIPOLAR COAGULATING CURVED SCISSORS FELL OFF. SURGEON WAS UNABLE TO RETRIEVE CERAMIC PIECE. MFR HAS NOTIFIED CO THAT THE COMPONENT THAT SEPARATED IS ZIRCONIA CERAMIC WHICH HAS BEEN TESTED TO BIOCOMPATIBILITY STANDARDS. IT WAS ALSO NOTED THAT THIS PRODUCT COMPL... | foitext1996.txt |
29207 | 16543 | D | 1 | null | THE DEVICE IS ON ALERT. | foitext1996.txt |
29216 | 21395 | D | 1 | null | PT HAS RECURRENT SUPERFICIAL INFECTIONS, STIMULATOR SITE 2 DEGREE OBESITY. STIMULATOR WAS FUNCTIONING WELL. | foitext1996.txt |
29228 | 17315 | D | 1 | null | ATRIAL LEAD IMPLANTED THEN EXPLANTED BECAUSE SCREW WOULD NOT EXTEND OR RETRACT. | foitext1996.txt |
29292 | 21463 | D | 1 | null | THE NEW MODEL'S SPRING DOES NOT HOLD AS LONG AS THE ORIGINAL MODEL IN ORDER TO GET GOOD PENETRATION NEEDED TO DRAW BLOOD. SHOULD BE GOOD FOR FIVE YEARS AND IT IS ONLY LASTING THIRTY DAYS. | foitext1996.txt |
29309 | 19850 | D | 1 | null | "ONCE STYLET INSERTED, UNABLE TO DRAW STYLET BACK. BALLOON WAS INFLATED AND DEFLATED AND NEITHER THE STYLET NOR CATHETER COULD BE REMOVED WITHOUT TRAUMA TO PT. UNABLE TO PASS CATHETER TO BLADDER. REPLACED WITH A SECOND CATHETER. BLEEDING WAS NOTED." UNUSED SAMPLES FROM THE SAME LOT AS ALLEGEDLY FAILED PRODUCT WILL BE R... | foitext1996.txt |
29229 | 21397 | D | 1 | null | ADD'L INFO PER CALL TO RPTR 1/3/95: THE LATEST IMPLANT IMPLANTED IN RPTR ON 12/4/95, IS ALSO LEAKING AND HAS BEEN TESTED SEVERAL TIMES SINCE IT WAS IMPLANTED IN RPTR. AS OF THIS DATE, THE LEAK CONTINUES. (ALSO SEE 1008050, 1008051). | foitext1996.txt |
29238 | 17317 | D | 1 | null | RIGHT SIDE ENCAPSULATED SHORTLY AFTER IMPLANT. | foitext1996.txt |
29253 | 20767089 | D | 1 | null | SPONGE FRAGMENTED IN WOUND. PT DID NOT REQUIRE MEDICAL TREATMENT. THERE WAS NO INJURY. | foitext1996.txt |
29255 | 54232 | D | 1 | null | CR BARD, INC. 8195 INDUSTRIAL BLVD, COVINGTON, GA 30209-2695. AN MDR INJURY REPORT WAS FILED ON THIS EVENT ON 1/30/96, ACCESS NUMBER M745505. | foitext1996.txt |
29256 | 54227 | D | 1 | null | BASED UPON THE INFO RECEIVED FROM THE CUSTOMER THE EVENT WAS JUDGED NOT TO BE REPORTABLE UNDER THE MDR RULES. | foitext1996.txt |
29295 | 17345 | D | 1 | null | BREAST IMPLANT RUPTURED LEADING TO DISFIGUREMENT AND SILICONE LEAK INTO BODY. MAMMOGRAM DID NOT SHOW RUPTURE, MRI OF CHEST SHOWED RUPTURE. REMOVED 8/24/94. SHE HAS HAD NO TREATMENT AT THIS TIME TO CLEAN BLOOD OR TISSUE OF SILICONE. | foitext1996.txt |
29307 | 20141217 | D | 1 | null | PT COMPLAINED OF PAIN. IT WAS FOUND THAT THE PLASTIC HAD SEPARATED FROM THE METAL. PATELLA WAS REPLACED WITH ANOTHER COMPONENT. | foitext1996.txt |
29328 | 16605 | D | 1 | null | ROD BROKE WHILE IN PT'S LEG. SURGERY TO REMOVE. MD STATED THIS WAS A STRESS BREAK. ADD'L CAT NUMBER 3517-3010. | foitext1996.txt |
29368 | 21502 | D | 1 | null | MD REPORTED VERBALLY THAT PT'S SHUNT HAD "FRACTURED," REQUIRING SURGICAL REVISION ON 11/3/95. MD GAVE IMPLANT TO SALES REP ON 11/9/95. | foitext1996.txt |
29385 | 16637 | D | 1 | null | PT WAS IN CARDIAC CATHETER LAB FOR A PROCEDURE. THE CARDIAC CATHETER BROKE OFF IN THE AORTA AND THE PT HAD TO GO TO THE OR FOR IT TO BE RETRIEVED. | foitext1996.txt |
29430 | 19937 | D | 1 | null | BAGS ARE NOT DRAINING AND URINE IS BACKING UP INTO THE PT'S BLADDER. PT DEVELOPED AN INFECTION AND WAS TREATED WITH ANTIBIOTICS. | foitext1996.txt |
29337 | 20689526 | D | 1 | null | RUPTURED IMPLANT CAUSED AUTO-IMMUNE DISORDER. PROGNOSIS IS POOR. PT HAS LIFE THREATENING PROBLEM. | foitext1996.txt |
29359 | 16609 | D | 1 | null | PT HAD A TOTAL HIP REPLACEMENT WITH PRODUCT ON 12/9/87. PT EXPERIENCED PAIN IN HIP AND IT WAS DISCOVERED THAT THE POLY ACETABULAR LINER HAD FAILED - LINER FRACTURED INTO SEVERAL PIECES. | foitext1996.txt |
29363 | 16366188 | D | 1 | null | C-ARM USED FOR PT IN ICU. WHEN MOVING C-ARM OUT OF ROOM FRONT WHEEL BOLT SNAPPED OFF. NO ONE INJURED. UNIT REMOVED AND WILL NOT BE PUT BACK INTO SVC AT THIS TIME. | foitext1996.txt |
29387 | 17396 | D | 1 | null | VENTILATOR CONNECTED TO PT BEGAN TO SMELL OF BURNING PLASTIC AND BEGAN TO OMIT SMOKE. THE SCREEN WENT BLACK BUT THE VENTILATOR CONTINUED TO FUNCTION UNTIL IT WAS DISCONNECTED FROM THE PT AND TAKEN OUT OF SERVICE. NO INJURY. | foitext1996.txt |
29388 | 21505 | D | 1 | null | PT'S CHEST AREA RECEIVED 1ST DEGREE BURNS DURING CARDIOVERSION USING THE PADS. | foitext1996.txt |
29396 | 15441618 | D | 1 | null | PT UNDERWENT AN ATTEMPTED LEFT ILIAC ANGIOPLASTY. PREVIOUSLY THE PT HAD STENTS AND A CATHETER, NINE FRENCH SHEATH WAS PASSED BEYOND THE STENT. UPON REMOVING THE SHEATH RESISTANCE WAS MET, THE CATHETER MANIPULATED AND AFTER REMOVING IT, THE END WAS RAGGED SUGGESTING THAT IT HAD TORN ON THE INDWELLING VASCULAR STENT WITH... | foitext1996.txt |
29407 | 91752 | D | 1 | null | EVALUATION FINDINGS: THIS BOWL WAS RETURNED FOR EVALUATION. CO WAS NOT ABLE TO REPLICATE THE NOISE OR THE LEAK REPORTED. CO DID FIND ABRASION SIGNS ON THE RIBS OF THE HEADER AND THE EFFLUENT TUBE, INDICATING THAT THE BOWL WAS NOT SEATED PROPERLY. THIS COULD HAVE CAUSED THE NOISE AND THE LEAK. | foitext1996.txt |
29418 | 19935 | D | 1 | null | DENTAL IMPLANT FAILED IN FUNCTION. | foitext1996.txt |
30108 | 17768 | D | 1 | null | LEAD WAS REMOVED FROM SERVICE BECAUSE "LEAD FRACTURE AT PIN COIL INTER-CONNECT."INVALID DATA - REGARDING SINGLE USE LABELING OF DEVICE. PATIENT MEDICAL STATUS PRIOR TO EVENT: INVALID DATA. THERE WAS NOT MULTIPLE PATIENT INVOLVEMENT.INVALID DATA - ON DEVICE SERVICE/MAINTENANCE. NO DATA - REGARDING DATE LAST SERVICED... | foitext1996.txt |
29449 | 17435 | D | 1 | null | DURING IMPLANTATION IT WAS NOTED THAT HAPTICS ON ONE SIDE OF THE LENS WERE BROKEN. THEREFORE THE LENS WAS NOT USED. SURGERY WAS ABORTED AS UNABLE TO IMPLANT A 2ND LENS. | foitext1996.txt |
29463 | 16649 | D | 1 | null | IMPLANT FAILED TO OSSEOINTEGRATE. | foitext1996.txt |
29488 | 19945 | D | 1 | null | STIMULATOR WAS NOT FUNCTIONING. | foitext1996.txt |
30114 | 20768014 | D | 1 | null | THE PT IS A 65-YR-OLD FEMALE WITH UNSTABLE ANGINA WHO WAS EN ROUTE FOR TRANSFER TO HOSP. SHE HAD A BALLOON PUMP IN PLACE. SUDDENLY, THE PUMP READ "AUTOFILL FAILURE" AND WOULD NOT FILL VIA THE MANUAL FILL BUTTON. THE HELIUM TANK WAS FOUND TO BE EMPTY AND A SLOW LEAK AT THE NOZZLE ERRUPTED. THE PT IMMEDIATELY COMPLAINED ... | foitext1996.txt |
29441 | 21965130 | D | 1 | null | SOURCE ADDED 500 CC OF NORMAL SALINE AND 250 ML OF TAXOL TO THE CONTAINER WHICH WAS THEN SPIKED WITH VENTED TUBING AND DELIVERED TO FLOOR. CONTAINER WAS PLACED ON THE COUNTERTOP. THE PT REPORTED A LOUD POP AND THE CONTAINER FELL TO THE FLOOR AND BROKE FURTHER. | foitext1996.txt |
29459 | 20767091 | D | 1 | null | CASSETTE ON PUMP DISCONNECTED ABNORMALLY. | foitext1996.txt |
29478 | 16651 | D | 1 | null | PIG TAIL BLOCKED "(AS NEW)" FLUID WILL NOT PASS. UNIT SET UP IN NEONATAL ICU. (LOOKS LIKE MFG PROBLEM.) | foitext1996.txt |
29487 | 16675 | D | 1 | null | DURING STEAM STERILIZATION, PROCESS INDICATOR TAPE BECOMES LOOSE AND ADHESIVE FAILS TO REMAIN ATTACHED TO PACKS IN THE AUTOCLAVE. ONLY THIS LOT APPEARS TO BE AFFECTED BY PROBLEM. TAPE FROM OTHER LOTS IS WORKING OK. | foitext1996.txt |
29486 | 21558 | D | 1 | null | CATHETER FRACTURED AND BROKE OFF EVEN WITH THE TOP OF THE PATIENT PORT, LEAVING APPROX 3/4 CM IMBEDDED IN THE PATIENT PORT. | foitext1996.txt |
29499 | 16197 | D | 1 | null | RPTR HAD BREAST IMPLANTS PUT IN FOLLOWING MASTECTOMY FOR FIBROCYSTIC DISEASE. SHE HAS HAD HER BREAST IMPLANTS REPLACED TWICE. SHE HAS HAD A TOTAL OF 3 BREAST SURGERIES ON EACH BREAST INCLUDING THE FIRST IMPLANT SURGERY. SHE HAD CALCIUM DEPOSITS. A MEDICAL PROFESSIONAL HAS CONFIRMED THE PRESENCE OF SILICONE IN THE LEFT ... | foitext1996.txt |
29521 | 19982 | D | 1 | null | OCCLUDED SHUNT. | foitext1996.txt |
29543 | 19986 | D | 1 | null | DURING CODE CENTRAL LINE SNAPPED WHILE SUTURE WAS BEING APPLIED. PT CODED TWICE AFTERWARDS AND EXPIRED. | foitext1996.txt |
29555 | 21061 | D | 1 | null | HIGH PEAK PRESSURE ALARM GOING OFF ON VENTILATOR. PT CHECKED AND SUCTIONED, ALARM CONTINUED. VENT CHECKED AND TUBE THAT ACTIVATED EXHALATION VALVE FOUND TO BE DISCONNECTED, THIS TUBING IS BEHIND ACCESS COVER AND NOT IN PLAIN VIEW. | foitext1996.txt |
30123 | 19770 | D | 1 | null | EXCHANGE BREAST IMPLANTS.INVALID DATA - REGARDING SINGLE USE LABELING OF DEVICE. PATIENT MEDICAL STATUS PRIOR TO EVENT: SATISFACTORY CONDITION. THERE WAS NOT MULTIPLE PATIENT INVOLVEMENT.INVALID DATA - ON DEVICE SERVICE/MAINTENANCE. NO DATA - REGARDING DATE LAST SERVICED. SERVICE PROVIDED BY: INVALID DATA. INVAL... | foitext1996.txt |
30137 | 21436 | D | 1 | null | TWO LEADS WERE REMOVED FROM SVC BECAUSE OF A "RATE SENSING LEAD, AND ANOTHER MODEL LEAD, WERE REMOVED FROM SVC BECAUSE OF A "RATE SENSING LEAD PROBLEM OF OVERSENSING AND INAPPROPRIATE SHOCKS." CPI ASSUMES THE LEADS WERE CAPPED AND REMAIN IMPLANTED. IMPLANTED 9/14/94. REMOVED FROM SVC APPROX 12/1/95. TOTAL IMPLANT TIME ... | foitext1996.txt |
29587 | 17510 | D | 1 | null | EXTRA-CORPOREAL MEMBRANE OXYGENATION PUMP STOPPED FUNCTIONING. MANUAL ROTATION WAS STARTED AS THE PROBLEM WAS IDENTIFIED. A NEW PUMP WAS BROUGHT TO THE BEDSIDE AND PUT INTO USE. OXYGEN SATURATIONS DROPPED INITIALLY AND THEN RETURNED TO NORMAL AS THE PROBLEM WAS CORRECTED. BIOMEDICAL ENGINEERING OPENED THE PUMP AND FOUN... | foitext1996.txt |
29565 | 16229 | D | 1 | null | CROSS-TALK BETWEEN TWO UNRELATED INTRA-CARDIAC ELECTROGRAM OUTPUTS CAUSING INCORRECT PERFORMANCE OF HEMODYNAMIC MONITORING SYSTEM DURING HEART CATHERIZARTION. | foitext1996.txt |
30133 | 19771 | D | 1 | null | THIS SUTURELESS MYOCARDIAL LEAD AND ANOTHER LEAD WERE REMOVED FROM SVC FOR INSULATION FAILURE. THE LEADS ARE CAPPED AND REMAIN IMPLANTED. IMPLANT DATE 7/29/93. OUT OF SVC DATE 9/7/95. TOTAL IN-SERVICE TIME 25 MONTHS.INVALID DATA - REGARDING SINGLE USE LABELING OF DEVICE. PATIENT MEDICAL STATUS PRIOR TO EVENT: INVALID... | foitext1996.txt |
30126 | 21701032 | D | 1 | null | EXCHANGE BREAST IMPLANTS WITH CAPSULECTOMY.INVALID DATA - REGARDING SINGLE USE LABELING OF DEVICE. PATIENT MEDICAL STATUS PRIOR TO EVENT: SATISFACTORY CONDITION. THERE WAS NOT MULTIPLE PATIENT INVOLVEMENT.INVALID DATA - ON DEVICE SERVICE/MAINTENANCE. NO DATA - REGARDING DATE LAST SERVICED. SERVICE PROVIDED BY: IN... | foitext1996.txt |
30129 | 16526 | D | 1 | null | DURING A ROUTINE REPLACEMENT OF AN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD), THE TWO (RATE SENSING) LEADS WERE REPAIRED AND REMAIN IMPLANTED. IMPLANTED 9/11/92 (IMPLANTED 39 MONTHS).INVALID DATA - REGARDING SINGLE USE LABELING OF DEVICE. PATIENT MEDICAL STATUS PRIOR TO EVENT: INVALID DATA. INVALID DATA - REGARDI... | foitext1996.txt |
30134 | 17908409 | D | 1 | null | THIS LEAD AND ANOTHER MODEL WERE "FOUND TO HAVE VISIBLE SENSING PROBLEMS - EROSION." BOTH LEADS WERE CAPPED AND REMAIN IMPLANTED. IMPLANTED 9/15/94. REMOVED FROM SVC 12/8/95. TOTAL IMPLANT TIME 14 MONTHS.INVALID DATA - REGARDING SINGLE USE LABELING OF DEVICE. PATIENT MEDICAL STATUS PRIOR TO EVENT: INVALID DATA. THER... | foitext1996.txt |
29604 | 20012 | D | 1 | null | DENTAL IMPLANT FAILED TO OSSEOINTEGRATE. | foitext1996.txt |
29651 | 20019 | D | 1 | null | DENTAL IMPLANT FAILED TO OSSEOINTEGRATE. | foitext1996.txt |
29666 | 16266 | D | 1 | null | NEEDLE ASSEMBLY CLOGGED (BLOCKED OR RESTRICTED) THUS LIMITING FLOW AND CAUSING AMBULATORY PUMPS TO ALARM WITH OCCLUSION. | foitext1996.txt |
29699 | 21127 | D | 1 | null | PT SUSTAINED V. FIB ARREST. IMPLANTED AICD DID NOT FUNCTION. ONCE PARAMEDICS ARRIVED AT SCENE, PT WAS DEFIBRILLATED SUCCESSFULLY. PT SUSTAINED ANOXIC BRAIN INJURY AND COMA. | foitext1996.txt |
29705 | 16274 | D | 1 | null | SURGEON WAS USING AN OSTEOTOME TO LOOSEN A FEMORAL KNEE PROSTHESIS INTRA-OPERATIVELY WHEN THE OSTEOTOME SLIPPED FROM THE SURGEON'S HAND, STRIKING HIS RIGHT PALM AND INFLICTING AN OPEN LACERATION TO THE BASE OF THE THUMB. AN ARTERIOLE WAS LACERATED WHICH WAS SUTURED BY ANOTHER SURGEON FOLLOWING IRRIGATION OF WOUND WITH ... | foitext1996.txt |
29711 | 16302 | D | 1 | null | HISTORY OF BILATERAL BREAST AUGMENTATION & SUBSEQUENT BILATERAL REDUCTION REPLACEMENT OF IMPLANTS. PT NOW HAS BILATERAL CAPSULAR CONTRACTURE AND PAIN; WANTS IMPLANTS REMOVED. | foitext1996.txt |
29630 | 20016 | D | 1 | null | EPIDURAL CATHETER WAS INSERTED IN LABORING OB PATIENT WITHOUT DIFFICULTY. WHEN CATHETER WAS REMOVED, IT BROKE, LEAVING APPROXIMATELY 12 CM. BOTH CATHETER AND WIRE REMAIN EMBEDDED IN PATIENT. PATIENT SUFFERED NO ADVERSE CONSEQUENCES. RECOMMENDATION FROM SEVERAL ANESTHESIOLOGISTS IS TO LEAVE THE PIECE OF DEVICE WHERE IT ... | foitext1996.txt |
29595 | 21086 | D | 1 | null | RPTR FINDS THAT CONICAL MEDICATION CUPS ARE NOT ACCURATE. THEY ARE OK FOR NON-SIGNIFICANT MEDICATIONS BUT HAVE A VARIED MARGIN OF ERROR FOR SIGNIFICANT MEDICATIONS - EITHER TOO LITTLE OR TOO MUCH. VARYING DEGREES OF OPACITY AND GRADUATION (FINE) OR (COARSE) ARE CONTRIBUTING FACTORS. RPTR SUGGESTS CYLINDRICAL MEASURING ... | foitext1996.txt |
29601 | 19167750 | D | 1 | null | DENTAL IMPLANT BROKE IN FUNCTION. | foitext1996.txt |
29608 | 21088 | D | 1 | null | DENTAL IMPLANT FAILED TO OSSEOINTEGRATE. | foitext1996.txt |
29632 | 17540 | D | 1 | null | DURING CANNULATION, IT WAS NOTICED THAT CANNULA WAS CRACKED NEAR THE CONNECTION TO THE CARDIOPULMONARY TUBING. THE LINE CLAMP WAS CLOSED. | foitext1996.txt |
29643 | 17542 | D | 1 | null | DENTAL IMPLANT FAILED TO OSSEOINTEGRATE. | foitext1996.txt |
29649 | 19892227 | D | 1 | null | DENTAL IMPLANT FAILED TO OSSEOINTEGRATE. | foitext1996.txt |
29664 | 17545 | D | 1 | null | WHILE EXPELLING AIR FROM LOW PRESSURE SYRINGE WITH INJECTOR, SYRINGE END EXPLODED OFF IN 2 PIECES, DUE TO INJECTOR PRESSURE. NO INJURY TO PT, OPERATOR OR EQUIPMENT. ERROR CODE ON THE INJECTOR REGISTERED E439. USER WAS INSTRUCTED TO CALL CO SERVICE NUMBER. | foitext1996.txt |
29669 | 21121 | D | 1 | null | EXTENSION SET WOULD NOT FLUSH, APPEARS TO HAVE DEFECT/BLOCKAGE. WAS NEVER ATTACHED TO PT. | foitext1996.txt |
29675 | 16268 | D | 1 | null | AFTER CORONARY INJECTIONS HAD BEEN COMPLETED, PT WENT INTO V-TACH AND WAS DEFIBRILLATED X2 @ 200 W/S WITH CONVERSION. PT REMAINED UNSTABLE, WITH REPEATED DETERIORATION OF RHYTHM AND SUBSEQUENT DEFIBRILLATIONS. PHYSICIAN INSERTED IABP CATHETER. ATTENDING RN'S STATED THAT IT WOULD NOT AUGMENT. GAS LINE WAS CHECKED FOR LE... | foitext1996.txt |
29721 | 20036 | D | 1 | null | DURING A LAPAROSCOPIC TUBAL STERILIZATION PROCEDURE, THE PT'S UTERUS WAS PERFORATED. PRODUCT INSTRUCTIONS REQUIRE SOUNDING OF THE UTERUS IN ORDER TO ACCURATELY CHOOSE APPROPRIATE SIZE DEVICE. UPON INVESTIGTION OF THE INCIDENT, IT IS BELIEVED THAT SOUNDING HAD NOT OCCURRED. THE TIP ITSELF WAS OBSERVED TO FUNCTION APPROP... | foitext1996.txt |
29727 | 20519266 | D | 1 | null | PERFORATOR DID NOT STOP AS SOON AS CRANIUM WAS PENETRATED. | foitext1996.txt |
29742 | 21187 | D | 1 | null | UPON ADMISSION, 1/6/96, THE PT HAD BEEN COMPLAINING OF RIGHT CHEST WALL PAIN AND RIGHT SHOULDER PAIN FOR THE LAST FOUR DAYS, BUT DENIED ANY PRODUCTIVE COUGH, DYSURIA, DIARRHEA, NAUSEA OR VOMITING. SHE WAS ADMITTED TO THE HOSP WITH NEUTROPENIC FEVER AND ON ADMISSION WAS FOUND TO BE ANEMIC. AFTER ADMISSION THE PT'S COURS... | foitext1996.txt |
29750 | 16309 | D | 1 | null | RPTR HAS RECEIVED TWO SEPARATE COMPLAINTS REGARDING THIS RPODUCT. 1) DIFFICULT TO TURN/CAN'T PRIME. TWO DEVICES RETURNED TO MFR. LOT #S UNKNOWN. 2) DEVICE INFUSING TOO FAST. SEVERAL INCIDENTS LATE 12/95, 3 INCIDENTS 11/29/96 (LOTS LISTED) 1 INCIDENT 1/25/96, 1 ON 1/26/96 (NO LOTS). DEVICES RETURNED TO MFR. | foitext1996.txt |
29781 | 17619 | D | 1 | null | A STANDARD OCCLUSION BALLOON CATHETER TESTED OUTSIDE THE PT AND THEN INSERTED INTO THE INTERNAL CAROTID ARTERY. THE BALLOON RUPTURED AND THE DEVICE WAS REMOVED. THE PT EXPERIENCED A NEUROLOGICAL EPISODE, WHICH COMPLETELY RESOLVED. THIRTY MINUTES LATER A SECOND BALLOON CATHETER WAS TESTED OUTSIDE THE PT AND INSERTED INT... | foitext1996.txt |
29809 | 17623 | D | 1 | null | RPTR HAS HAD NUMEROUS COMPLAINTS FROM CUSTOMERS THAT BOTH THEIR PHYSICIANS AND THE HOSP LAB HAVE CHECKED THE STRIPS AND FOUND THEM NOT TO BE ACCURATE. | foitext1996.txt |
29868 | 21268 | D | 1 | null | REMOVAL OF BILATERAL BREAST IMPLANTS THAT HAD RUPTURED INTO HUMAN CAPSULE. | foitext1996.txt |
29879 | 21270 | D | 1 | null | RIGHT MAMMARY SILICON FILLED IMPLANT RUPTURED. PT EXPERIENCED PAIN AND NEW ONSET OF FIBROMYALGIA, WHICH BELIEVED TO BE RELATED TO THE RUPTURED IMPLANTS. SURGERY REQUIRED TO EXPLANT THIS MAMMARY IMPLANT. | foitext1996.txt |
29749 | 21397491 | D | 1 | null | PT HAD A LEFT INGUINAL HERNIA REPAIR WITH MESH PLACEMENT ON 8/3/95. THE PT HAD INCREASING PAIN AND ON 9/1/95 HAD THE MESH REMOVED. AT THE TIME OF MESH REMOVAL IT WAS NOTED THAT THE PT HAD SOME PURULENT MATERIAL LOCATED IN THE AREA OF THE MESH ONLY. THE PT HAD NO VISUAL SIGNS OF INFECTION. CULTURES TAKEN OF THE PURULENT... | foitext1996.txt |
29753 | 17483373 | D | 1 | null | AN 89-YR-OLD FEMALE PT GOT OUT OF BED WITHOUT CALLING FOR ASSISTANCE, FELL, AND SUSTAINED A TRANSCERVICAL FRACTURE OF THE RIGHT PROXIMAL FEMUR, WITH DISPLACEMENT AND ANGULATION. THE PT UNDERWENT RIGHT HEMIARTHROPLASTY ON 12/12/95. THE BED ALARM FAILED TO ALARM. | foitext1996.txt |
29773 | 19077284 | D | 1 | null | A 74 YR OLD IN E.R. WITH CHEST PAIN. INITIALLY GIVEN SL NTG WITH RELIEF. WAS BEING MONITORED ON CARDIAC MONITOR. PT WENT INTO V-FIB AND ARRESTED. CARDIAC MONITOR DID NOT ALARM BUT SITUATION DETECTED BY STAFF IMMEDIATELY. THE PT WAS SUCCESSFULLY RESUSCITATED/CARDIOVERTED. DIAGNOSED WITH AN MI. HOSPITALIZED FOR 7 DAYS AN... | foitext1996.txt |
29777 | 16313 | D | 1 | null | DURING "CODE" DEFIBRILLATOR DID NOT DISCHARGE WHEN USED AT 20 & 50 JOULES. | foitext1996.txt |
29779 | 19575 | D | 1 | null | PT HAD EPIDURAL CATHETER IN PLACE FOR PAIN MANAGEMENT. UPON REMOVAL, DR NOTED TIP OF CATHETER APPEARED TO BE MISSING. X-RAY ORDERED SHOWS TIP IN SUBCUTANEOUS TISSUE. PT. SCHEDULED FOR LOCAL (SURGERY) FOR REMOVAL. NO COMPLICATIONS. PT TOLERATED PROCEDURE. DISCHARGED TO HOME. | foitext1996.txt |
29785 | 18338805 | D | 1 | null | THE CAP WAS NOT INTACT; THE CATHETER POPPED OUT OF THE PACKAGE DURING OPENING AND STUCK THE PT. CO WAS NOTIFIED OF THE INCIDENT. | foitext1996.txt |
29787 | 17765053 | D | 1 | null | A 9 1/2 PIECE OF CATHETER GUIDE WIRE DISCOVERED IN PTS ARTERY ON ROUTINE X-RAY. CATHETER HAD BEEN INSERTED INTO PT ON 7/18/95. ON 7/23/95 WIRE WAS REMOVED DURING RADIOLOGIC PROCEDURE USING A CATHETER AND A SNARE. | foitext1996.txt |
29794 | 20995661 | D | 1 | null | RPTR RECENTLY EXPERIENCED A SEVERE RESPIRATORY INFECTION AND HIS PHYSICIAN PUT HIM ON HOME OXYGEN. THE PLASTIC TUBES FURNISHED WITH THE EQUIPMENT TO DELIVER THE OXYGEN (BY NASAL CANNULA) CAUSED SEVERE IRRITATION TO HIS BRONCHIAL TUBES, HIS UPPER LIP AND HIS MOUTH. IT RESULTED IN SEVERE ASTHMATIC PROBLEMS AND HE WAS FOR... | foitext1996.txt |
29832 | 19583 | D | 1 | null | DENTAL IMPLANT FAILED TO OSSEOINTEGRATE. | foitext1996.txt |
29837 | 19584 | D | 1 | null | ADMITTING DIAGNOSIS HIATAL HERNIA. PERFORMED LAP NISSAN ON 1/16/96. PT RETURNED TO SURGERY ON 1/18/96 FOR POST-OP BLEEDING. PT SENT TO CCU AND EXPIRED ON 1/19/96. THE SCALPEL WAS USED DURING THE LAP NISSAN AND ACCORDING TO PHYSICIAN, WORKED AS EXPECTED. HOWEVER, IT WAS SINGLED OUT FOR REVIEW BECAUSE IT WAS NEW AND BECA... | foitext1996.txt |
29855 | 19587 | D | 1 | null | PT WAS BROUGHT TO ER AM OF 12/31/95 COMPLAINING OF HEADACHE AND N/V SHUNT SERIES OF X-RAYS WERE TAKEN AND IT WAS NOTED BY NEUROSURGEON THAT THE PLASTIC END OF SHUNT GOING INTO THE PUMP PART OF THE VP SHUNT ASSEMBLY HAD BROKEN AND SEPARATED, LOSING ITS CONTINUITY. | foitext1996.txt |
29872 | 18447021 | D | 1 | null | AFTER PT HAD BEEN ON DIALYSIS FOR 1 HR, WHEN MAKING ROUTINE DIALYSIS UNIT ROUNDS, THE DIALYSIS CARE GIVER NOTED A SMALL AMOUNT OF DRIED BLOOD ALONG THE TOP OF THE HEADER OF THE HEMODIALYZER BEING USED FOR THAT TREATMENT. DIALYSIS WAS DISCONTINUED AND RESTARTED WITH NEW DIALYZER AND BLOOD TUBING. PHYSICIAN WAS NOTIFIED.... | foitext1996.txt |
29874 | 21269 | D | 1 | null | PT WENT FROM MULTIPLE PVCS TO TORSADE V-TACH AND THEN V-FIB. MONITOR FAILED TO ALARM FOR V-TACH OR V-FIB. PT REQUIRED DEFIBRILLATION TO RESTORE NORMAL RHYTHM. | foitext1996.txt |
30158 | 21473 | D | 1 | null | THE BUTTON ON INJECTOR CORD STUCK CAUSING THE PT TO RECEIVE MULTIPLE INJECTIONS OF DYE. THE PT FIBRILLATED AND REQUIRED DEFIBRILLATION. NO PERMANENT INJURY.DEVICE NOT LABELED FOR SINGLE USE. PATIENT MEDICAL STATUS PRIOR TO EVENT: SATISFACTORY CONDITION. THERE WAS MULTIPLE PATIENT INVOLVEMENT. NUMBER OF PATIENTS INV... | foitext1996.txt |
29892 | 21272 | D | 1 | null | PT HAD PACEMAKER IMPLANTED 1/16/96. TO OR ON 1/19/96 FOR EROSION OF PACEMAKER POCKET. PT RETURNS TO SURGERY TODAY FOR SAME THING. | foitext1996.txt |
29905 | 21274 | D | 1 | null | DURING A LAP CHOLECYSTECTOMY, CLIPS FELL OFF OF THE CLIP APPLIER BEFORE FORMING. ALL CLIPS RETRIEVED. ANOTHER CLIP APPLIER MISFIRED. MFR'S PRODUCT REP NOTIFIED. | foitext1996.txt |
29969 | 19673 | D | 1 | null | WHILE USING THE CLIP APPLIER, IT MALFUNCTIONED AND WOULD NOT RELOAD A CLIP. CARTRIDGE WAS THE ORIGINAL THAT CAME WITH APPLIER. | foitext1996.txt |
30164 | 16577 | D | 1 | null | A PT WITH AN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND TWO (RATE SENSING) LEADS EXPERIENCED OVERSENSING/RECEIVED A DIVERTED SHOCK IN NORMAL SINUS RHYTHM DURING A FOLLOW-UP TEST PROCEDURE. AN INVASIVE PROCEDURE WAS PERFORMED AND THE PHYSICIAN ELECTED TO REPLACE THE TWO (RATE SENSING) LEADS. BOTH LEADS IMPLANTED 7... | foitext1996.txt |
30171 | 17369 | D | 1 | null | THE WRENCH STRIPPED WHILE TIGHTENING THE NUTS ON THE CLAMP.INVALID DATA - REGARDING SINGLE USE LABELING OF DEVICE. PATIENT MEDICAL STATUS PRIOR TO EVENT: INVALID DATA. THERE WAS NOT MULTIPLE PATIENT INVOLVEMENT.INVALID DATA - ON DEVICE SERVICE/MAINTENANCE. NO DATA - REGARDING DATE LAST SERVICED. SERVICE PROVIDED B... | foitext1996.txt |
29906 | 52748 | D | 1 | null | ASPEN LABS, 14603 E FREMONT AVE, ENGLEWOOD, CO 80112. THE ELECTROSURGERY ACCESSORY RETURNED TO ASPEN LABS ON THE ABOVE REFERENCED RA WAS LOGGEN AS A COMPLAINT INVESTIGATION WITH THE INDICATED NUMBER. THE INVESTIGATION HAS BEEN COMPLETED AND THE DEVICE WAS DISPOSITIONED AS DESCRIBED BELOW. THIS LETTER IS PROVIDED TO INF... | foitext1996.txt |