--- license: cc-by-4.0 tags: - mri - qsm - quantitative-susceptibility-mapping - neuroimaging pretty_name: QSMbly example data --- # QSMbly example data One in-vivo subject, used as the "Load example data" dataset in [QSMbly](https://qsmbly.neurodesk.org/) and fetched by the browser at runtime. A single 5-echo 3D gradient-echo acquisition at 1 mm isotropic, acquired on a Siemens MAGNETOM Prisma Fit. Magnitude and phase are split one file per echo, since QSMbly's loader reads 3D volumes and derives the echo count from the file list. | | | | --- | --- | | Matrix | 184 × 256 × 128, 1 mm isotropic | | Echoes | 5 — TE 5 / 11 / 17 / 23 / 29 ms | | TR / flip | 35 ms / 15° | | Sequence | `gre_bridge_1mm_psn_adapt` (product GRE, GRAPPA R=2) | | Scanner | MAGNETOM Prisma Fit (XR VA30A) | | Phase | radians, wrapped | ## Field strength `MagneticFieldStrength` is **2.8946 T**, the accurate field derived from the scanner's carrier frequency (`ImagingFrequency` 123.243444 MHz ÷ 42.576384 MHz/T), not the nominal 3.0 T printed on the magnet. QSM scales ppm by this value, so the accurate one is the correct one to use. ## Echo times Recovered from the exam-card PDFs shipped with the acquisition and, for the Pulseq consensus sequence in the wider study, from [HarmonizedMRI/megre_label](https://github.com/HarmonizedMRI/megre_label). They are not the DICOM sidecar values, which carry placeholder timings for some protocols. ## Provenance Packed from the 2026-08-20 QSM harmonization acquisition (MGH bays 4/5) — one subject scanned on two Siemens 3T scanners under four protocols × 3 runs. The full study is on OSF at [osf.io/gkemr](https://osf.io/gkemr/); this repo holds the `prisma-bridge-run1` acquisition only, re-split into per-echo files. The same acquisition is available to QSMxT users via `qsmxt example`.