| <h1 style="text-align:center">CHRIS: <font color="#0070C0">CH</font>emical <font color="#0070C0">RIS</font>k calculator - Bulk leachables</h1> | |
| </header> | |
| <h2 id="context-of-use-cou">Context of Use (COU)</h2> | |
| <p>The bulk leachable module of the CHemical RISk calculator (CHRIS) is | |
| intended to conduct screening level risk assessments to aid in the | |
| biocompatibility evaluation of bulk additives and impurities in | |
| polymeric medical device components. These assessments can assist device | |
| manufacturers by providing instantaneous feedback on whether the | |
| presence of the bulk chemical would require additional justification | |
| and/or testing to demonstrate acceptable biological risk. The output of | |
| CHRIS is a conservative margin of safety (MOS = toxicological safety | |
| limit ÷ exposure dose) value for a bulk chemical contained within a | |
| polymeric medical device component. Based on the MOS value, the | |
| calculator determines if further assessment of one or more | |
| biocompatibility endpoints is necessary for the specific chemical.</p> | |
| <p>Because CHRIS only addresses compounds with a distribution that is | |
| macroscopically homogeneous within the matrix, the tool can only be used | |
| to assess bulk additives and impurities. Therefore, only compounds that | |
| are introduced either intentionally or unintentionally during synthesis | |
| (e.g., residual monomers and oligomers, catalysts, initiators) or | |
| compounding (e.g., stabilizers, antioxidants, plasticizers) are within | |
| scope. Surface residuals from processing, cleaning, and sterilization | |
| are excluded. Also, CHRIS requires the total amount of the chemical to | |
| be established in advance, e.g., based on a certificate of analysis. | |
| Further CHRIS only addresses individual chemicals; therefore, a | |
| favorable outcome by CHRIS does not imply acceptable biological risk for | |
| the final finished form of a medical device. CHRIS is also not intended | |
| to establish device classification or identify biocompatibility | |
| requirements.</p> | |
| <p>CHRIS provides clinically relevant, yet still conservative, exposure | |
| dose estimates using a physics-based transport model for polymeric | |
| systems where transport data are available to support the use of the | |
| model. The model applies worst-case boundary conditions for release of a | |
| substance from the polymer matrix and is based on five (5) primary | |
| assumptions:</p> | |
| <ol type="1"> | |
| <li>The clinical use environment does not cause the polymer matrix to | |
| swell or degrade.</li> | |
| <li>Manufacturing processes do not impact the stability of the | |
| polymer.</li> | |
| <li>The chemical is homogeneously distributed throughout the | |
| polymer.</li> | |
| <li>The total amount of the chemical is present in dilute concentrations | |
| (<= 2 m/v %).</li> | |
| <li>Any particles/aggregates of the chemical present in the polymer are | |
| much smaller than the smallest component dimension (<= 50x).</li> | |
| </ol> | |
| <p>While these assumptions are typically valid for bulk additives and | |
| impurities in biostable polymers, users of CHRIS must confirm | |
| conformance to the underlying assumptions or provide supporting | |
| justification to ensure compliance for a given system. Further, CHRIS | |
| only enables system specific exposure estimates for fifty-three (53) | |
| polymeric systems that are generally biostable (non-swelling and | |
| non-degrading). These polymers are listed below. To estimate chemical | |
| release based on the model, the diffusion coefficient of the chemical in | |
| the polymer matrix must be specified. For the fifty-three (53) listed | |
| polymeric systems, a worst-case (upper bound) diffusion coefficient, as | |
| a function of molecular weight, has been established based on data from | |
| the literature. For polymer matrices that are not included in this list, | |
| CHRIS assigns an ultra-conservative diffusion coefficient that assumes | |
| the polymer has the properties of water. Note that the worst-case | |
| diffusion coefficient is only defined over a molecular weight range of | |
| up to 1100 g/mol. Therefore, for substances with a molecular weight > | |
| 1100 g/mol, the value of the diffusion coefficient assuming a molecular | |
| weight of 1100 g/mol can be used as a conservative value.</p> | |
| <p>In the absence of adequate toxicological and exposure data for a | |
| chemical in a polymeric matrix, a toxicological risk assessment can be | |
| conducted for systemic biocompatibility endpoints by comparing the | |
| exposure estimate to an appropriate threshold of toxicological concern | |
| (TTC). This is the approach used by CHRIS in this module. The TTC values | |
| are based on systemic toxicity, thus CHRIS can address acute systemic | |
| toxicity, subacute/subchronic toxicity, genotoxicity, carcinogenicity, | |
| and reproductive and developmental toxicity. It does not, however, | |
| address cytotoxicity, sensitization, irritation, hemocompatibility, | |
| material mediated pyrogenicity, or implantation. Therefore, an MOS >= | |
| 1 implies the chemical will not raise a safety concern with respect to | |
| only the systemic biocompatibility endpoints, provided the chemical is | |
| not within the cohort of concern, which is reflected in the output of | |
| CHRIS.</p> | |