Main Protocol

Performer: Hamza

Purpose / Hypothesis

Whether pulse can be detected. Whether distance to tube can be estimated basen on signal.


Equipment and Setup

  • Needle size: 22G
  • Needle tip: Quincke
  • Used sensor: M.4.4.0.22G
  • Punctured material: Candle gel

Phantom setup:
A fully filled tube embedded inside a candle gel box. We needed needle tip distances relative to the tube, not the surface of the material, in order to compare the distance for level measurements.
For this purpose, a candle gel box with an embedded tube was built.

alt text


Experimental Setup

  • Tube size: 10 × 8 mm
  • Pump speed: 22 rpm
  • Measurements performed at room temperature

Distances

  • Vertical distance from tube center to bottom of box: 3.5 cm
  • Distance from tube center to top surface: 6.3 cm
  • Candle gel surface begins at a radial distance of 6.3 cm from the tube center.

Measurement Points

All measurement points are defined relative to the tube center.

  • P2 is the outermost point (needle tip 5 mm inside the gel).
  • Located 5.8 cm from the tube center.
  • Each subsequent point moves toward the tube center with 5 mm (0.5 cm) spacing.

During tube recording the needle was moved until 8 mm distance from the tube surface, which is why recordings exist only up to P12.

Radial Distances

Point Distance from Tube Center Total Needle Shaft Inside
P2 5.8 cm 5 mm
P3 5.3 cm 10 mm
P4 4.8 cm 15 mm
P5 4.3 cm 20 mm
P6 3.8 cm 25 mm
P7 3.3 cm 30 mm
P8 2.8 cm 35 mm
P9 2.3 cm 40 mm
P10 1.8 cm 45 mm
P11 1.3 cm 50 mm
P12 0.8 cm 55 mm
P13 0.3 cm 60 mm
P14 −0.2 cm 65 mm
P15 −0.7 cm 70 mm
P16 −1.2 cm 75 mm
P17 −1.7 cm 80 mm

Each consecutive point is separated by 5 mm along the radial direction toward the tube center.

The distance to the tube in the Z axis in the recorded signal can be taken from the filename (see Filename Meaning section).


Surface Measurement Pattern

Measurements were performed on a grid placed on the candle gel surface.

Grid Layout

  • 3 columns: j, h, f
  • Each column contained 11 rows (10 rows + tube row)

The central row corresponds to the row where the tube is located:

  • 5 rows above the tube
  • 5 rows below the tube

Row Spacing

  • 10 mm spacing between most adjacent rows (1–5, 6–10)
  • 15 mm spacing between:
    • row 5 and tube
    • tube and row 6

alt text


Filename Meaning

Example filename:
h1_5.8_candlegel_22g_quincke_M.4.4.0.22G_2026-03-03_16.28.39.wav \

Format:
column-row_distanceToTubeInCMS_material_needle-gauge_tip_sensor-type_timestamp

distanceToTubeInCM/S

  • Measured in the Z axis
  • Independent distance originating from the row

Additional notes:

  • Sensor placed on channel 0
  • Channel 1 not connected

Needle Tracking Protocol

  • Marker size: 4 cm
  • Marker border size: 1 cm

Offsets:

  • Needle offset: 9.2 cm
  • Z-offset: 0.95 cm

Distances:

  • Tube center to table surface: 50 mm
  • Tube center to gel surface: 63 mm
  • Gel surface to table: 114 mm