Main Protocol
Performer: Hamza
Purpose / Hypothesis
Whether pulse can be detected. Whether distance to tube can be estimated basen on signal.
Equipment and Setup
- Needle size: 22G
- Needle tip: Quincke
- Used sensor: M.4.4.0.22G
- Punctured material: Candle gel
Phantom setup:
A fully filled tube embedded inside a candle gel box.
We needed needle tip distances relative to the tube, not the surface of the material, in order to compare the distance for level measurements.
For this purpose, a candle gel box with an embedded tube was built.
Experimental Setup
- Tube size: 10 × 8 mm
- Pump speed: 22 rpm
- Measurements performed at room temperature
Distances
- Vertical distance from tube center to bottom of box: 3.5 cm
- Distance from tube center to top surface: 6.3 cm
- Candle gel surface begins at a radial distance of 6.3 cm from the tube center.
Measurement Points
All measurement points are defined relative to the tube center.
- P2 is the outermost point (needle tip 5 mm inside the gel).
- Located 5.8 cm from the tube center.
- Each subsequent point moves toward the tube center with 5 mm (0.5 cm) spacing.
During tube recording the needle was moved until 8 mm distance from the tube surface, which is why recordings exist only up to P12.
Radial Distances
| Point | Distance from Tube Center | Total Needle Shaft Inside |
|---|---|---|
| P2 | 5.8 cm | 5 mm |
| P3 | 5.3 cm | 10 mm |
| P4 | 4.8 cm | 15 mm |
| P5 | 4.3 cm | 20 mm |
| P6 | 3.8 cm | 25 mm |
| P7 | 3.3 cm | 30 mm |
| P8 | 2.8 cm | 35 mm |
| P9 | 2.3 cm | 40 mm |
| P10 | 1.8 cm | 45 mm |
| P11 | 1.3 cm | 50 mm |
| P12 | 0.8 cm | 55 mm |
| P13 | 0.3 cm | 60 mm |
| P14 | −0.2 cm | 65 mm |
| P15 | −0.7 cm | 70 mm |
| P16 | −1.2 cm | 75 mm |
| P17 | −1.7 cm | 80 mm |
Each consecutive point is separated by 5 mm along the radial direction toward the tube center.
The distance to the tube in the Z axis in the recorded signal can be taken from the filename (see Filename Meaning section).
Surface Measurement Pattern
Measurements were performed on a grid placed on the candle gel surface.
Grid Layout
- 3 columns:
j,h,f - Each column contained 11 rows (10 rows + tube row)
The central row corresponds to the row where the tube is located:
- 5 rows above the tube
- 5 rows below the tube
Row Spacing
- 10 mm spacing between most adjacent rows (
1–5,6–10) - 15 mm spacing between:
- row 5 and tube
- tube and row 6
Filename Meaning
Example filename:
h1_5.8_candlegel_22g_quincke_M.4.4.0.22G_2026-03-03_16.28.39.wav \
Format:
column-row_distanceToTubeInCMS_material_needle-gauge_tip_sensor-type_timestamp
distanceToTubeInCM/S
- Measured in the Z axis
- Independent distance originating from the row
Additional notes:
- Sensor placed on channel 0
- Channel 1 not connected
Needle Tracking Protocol
- Marker size: 4 cm
- Marker border size: 1 cm
Offsets:
- Needle offset: 9.2 cm
- Z-offset: 0.95 cm
Distances:
- Tube center to table surface: 50 mm
- Tube center to gel surface: 63 mm
- Gel surface to table: 114 mm