- l Input: FASTQ |
- l Output: BAM, MEX, CSV, HDF5, Web Summary, .cloupe/.loupe |
## Loupe Browser |
Loupe Browser is an interactive data visualization application that requires no prior programming knowledge. |
- l Input: .cloupe, .vloupe |
- l Output: Data visualization, including t-SNE and UMAP projections, custom clusters, differentially expressed genes |
- l Operating System: MacOS, Windows |
Tab 1 Text |
Tab 2 Text |
Tab 3 Text |
Tab 4 Text |
Tab 5 Text |
Tab 6 Text |
<!-- image --> |
## Troubleshooting |
<!-- image --> |
| GEM Generation & Barcoding | 67 | |
|------------------------------|------| |
71 |
## GEM Generation & Barcoding |
STEP |
1.2 |
Load Chromium GEM-X Chip |
## 1.4d |
After Chip is removed from the instrument and the wells are exposed. |
Normal |
<!-- image --> |
Gasket holes are aligned with the sample and gel bead wells. |
All recovery wells are similar in volume and opacity. |
<!-- image --> |
## Impacted |
<!-- image --> |
Gasket holes are misaligned with the gel bead wells. Open and close the chip holder slowly once. |
One of the rows of Chromium X/iX gasket will not align with the GEM-X chip. This is normal and will not impact the assay. |
<!-- image --> |
Recovery well A indicates a wetting failure. Not all wetting failures may present themselves with excess bubbles (foam). |
<!-- image --> |
Recovery wells A-C show normal emulsions. Not all normal emulsions may have small (in size and number) bubbles in the recovery wells. |
Inspecting emulsions in pipette tips is recommended for diagnosing emulsion failures. |
Tab 1 Text |
Tab 2 Text |
Tab 3 Text |
Tab 4 Text |
Tab 5 Text |
Tab 6 Text |
## STEP |
1.4e Transfer GEMs |
from Chip All liquid levels are similar in volume and opacity without air trapped in the pipette tips. |
<!-- image --> |
## Normal |
<!-- image --> |
<!-- image --> |