Virginia,
I'm not sure if this applicable but ours sometimes get stuck in the up
position. if the stylus is raised to high, about 2" above the surface, then
it might have jammed the Z axis lead screw in the top position. You have to
take off the monitor and the top cover ( I actually just drilled a hole
through the top cover roughly near the top of the Z axis removing the cover
seemed to much hassle at the time) and use a large flat head screw driver
in the end of the lead screw, to break it loose and drive it back down. As I
precaution I train my students not raise the stylus more then about 12 to
18mm above the chuck surface.
Also, our unit sometimes will get stuck in the down position, and then the
system becomes unresponsive. I've just assumed that this is some glitch in
the software, because if power cycle the system, it seems to correct itself,
and the works normally afterwards. the unit is a 1991 vintage tool and
obsolete by most standards so I've just learned to live with it, until money
comes along for a newer model.
Richard Everly
University of South Florida
College of Engineering
Nanotechnology Research and Education Center
4202 E. Fowler Ave.
MS: ENG 030
Tampa, FL 33620
Office: (813) 974-5365
Fax: (813) 974-3610
Cell: (813) 760-3934
http://www.nrec.usf.edu/
everly at usf.edu
-----Original Message-----
From: labnetwork <labnetwork-bounces at mtl.mit.edu> On Behalf Of
vsoares at inesc-mn.pt
Sent: Wednesday, April 5, 2023 1:39 PM
To: labnetwork at mtl.mit.edu
Subject: [labnetwork] Dektak 3030 ST stylus issues
[You don't often get email from vsoares at inesc-mn.pt. Learn why this is
important at https://aka.ms/LearnAboutSenderIdentification ]
Dear all,
our lab still has a Dektak 3030ST profilometer and we are having issues with
the stylus. First it would go down and wouldn't go up, then I managed to get
it working for a few measurements and know we have the opposite problem: the
stylus won't go down.
Plus I moved the optics to see if the sytlus would go up (or down, now that
it is in an upper position), but the scan head doesn't move at all.
Does naybody have any idea what this could be due to?
Thanks
Virginia
--
Melhores cumprimentos/Best Regards
Virginia Soares
http://www.inesc-mn.pt/
INESC-MN
Rua Alves Redol,9
1000-029 LISBOA
PORTUGAL
_______________________________________________
labnetwork mailing list
labnetwork at mtl.mit.edu
https://mtl.mit.edu/mailman/listinfo.cgi/labnetwork
[EXTERNAL EMAIL] DO NOT CLICK links or attachments unless you recognize the
sender and know the content is safe.
From vk409 at soe.rutgers.edu Tue Apr 11 11:40:56 2023
From: vk409 at soe.rutgers.edu (Vibhor Kumar)
Date: Tue, 11 Apr 2023 15:40:56 +0000
Subject: [labnetwork] Suggestions on fume hood
Message-ID: <BYAPR14MB32213A452FD034A573961C31A79A9@BYAPR14MB3221.namprd14.prod.outlook.com>
Dear Labnetwork Community,
We have a user who wants to access fume hood for spin coating of solution containing:
lanthanum nitrate hexahydrate (La(NO3)3?6H2O) (solid) + nickel(II) acetate tetrahydrate (Ni(CH3COO)2? 4H2O) (solid)
Mixed with
mixture of 2-methoxy ethanol and acetic acid (volume ratio = 9:1).
The user will be using his own spinner.
Our concerns are that :
1. Which Fume hood will be ideal for them. i.e., Acid OR Solvent?
2. Is there any contamination issue that affects other users OR disposal issue of used aluminum foils?
Thanks in advance,
Sincerely,
-Vibhor Kumar-
Research Scientist,
School of Engineering,
Rutgers, The State University of New Jersey, NJ
USA.
-------------- next part --------------
An HTML attachment was scrubbed...
URL: <https://mtl.mit.edu/pipermail/labnetwork/attachments/20230411/3145128d/attachment.html>
From lino.eugene at uwaterloo.ca Thu Apr 13 16:37:49 2023
From: lino.eugene at uwaterloo.ca (Lino Eugene)
Date: Thu, 13 Apr 2023 20:37:49 +0000
Subject: [labnetwork] Evaluation of heavy metal contamination in oxidation