> We use WIN95, 98, and XP....whatever the manufacturers were using when they
> wrote the operating software.
>
> cordially,
>
> Keith Bradshaw
> University of Texas at Dallas
--
Best regards,
P. Scott Harris, P.Eng.
H&L Associates
21 Parkmount Crescent
Nepean, Ontario K2H 5T3 Canada
From vito.logiudice at uwaterloo.ca Fri Oct 15 14:59:08 2010
From: vito.logiudice at uwaterloo.ca (Vito Logiudice)
Date: Fri, 15 Oct 2010 14:59:08 -0400
Subject: [labnetwork] ICP RIE etch of NbTiN: chamber contamination issues?
Message-ID: <4998D909302C254F82125395DA58AF780338DBBF11@CONNMAIL.connect.uwaterloo.ca>
Hello all,
We are contemplating the possibility of etching NbTiN in our ICP RIE chlorine etch system (Oxford Plasmalab 100 with ICP380 source). Initial feedback regarding potential chamber contamination issues are encouraging but we would appreciate any input from people whom may have had experience etching this material in the p...
Regards,
Vito Logiudice
Vito Logiudice P.Eng.
Director of Operations
QNC & RAC Fabrication Facilities
University of Waterloo
200 University Avenue West
Waterloo, ON
Canada N2L 3G1
Tel: 1-519-888-4567 ext. 38703
Fax: 1-519-888-7610
Cel: 1-519-500-2538
Email: vlogiudi at uwaterloo.ca
-------------- next part --------------
An HTML attachment was scrubbed...
URL: <https://mtl.mit.edu/pipermail/labnetwork/attachments/20101015/350fb8cb/attachment.html>
From hughes at illinois.edu Fri Oct 15 17:27:12 2010
From: hughes at illinois.edu (Hughes, John S)
Date: Fri, 15 Oct 2010 16:27:12 -0500
Subject: [labnetwork] System backups
In-Reply-To: <4CADEDCF.2020103@umn.edu>
References: <AANLkTimMtcFQQb8hF616FKEGhqtRi_Y1HbEHC_pmu_u8@mail.gmail.com>
<4CADEDCF.2020103@umn.edu>
Message-ID: <87BCD977-C874-464D-AF59-2A1375E8F0D0@illinois.edu>
Hello Tony,
As per your suggestion, I asked our IT folks to look into getting SSDs to use as backups for a lot of our process tool computers. We have quite a few "legacy" systems, and the question they asked me was: where can we get an SSD drive with an IDE connection and <2GB capacity, since some of our older systems can't handle...
If you, or anyone else, knows of a good vendor for such drives, I'd appreciate hearing about them.
Thanks
John
-------------------------------------------------------------
John S. Hughes Office: (217) 333-4674
Associate Director FAX: (217) 244-6375
Laboratory Operations hughes at illinois.edu<mailto:hughes at illinois.edu>
Micro and Nanotechnology Laboratory
University of Illinois at Urbana-Champaign
3114 Micro and Nanotechnology Laboratory
208 North Wright Street
Urbana, Illinois 61801 http://mntl.illinois.edu
-------------------------------------------------------------
On Oct 7, 2010, at 10:57 AM, tony whipple wrote:
Hello Keith;
Great topic.
We had similar concerns. Some of our equipment use DOS
and the computers that came with the system could only accept
a limited size of hard drive, which you can not buy anymore.
When a hard drive failed we tried buying old recycled hard drives but
found they did not last too long. And we could not replace all the old
computers with a new one since some systems might need an ISA slot.
The solution we found was to use smaller solid state drives.
That way we could keep our legacy computer. This has worked
so well that we make back ups of the whole drive and copy it on
to another solid state drive. This spare solid state drive can then
be swapped out when the original hard drive fails. Minimum
equipment down time. Only thing is that the back up work needs
to be done before you have problems. Making regular backups
from time to time helps, that way you do not lose to most current
recipes.
They make them in many sizes and configurations, IDE, SATA, and USB.
For backing up the drive I like to use an IDE to USB adapter and
plug that into the normal desk computer. From there you can
mirror or ghost the drive, or even store it on a remote server and
then copy it to the new spare solid state drive.
Regards, Tony W.
On 10/6/2010 2:30 PM, Keith Bradshaw wrote:
We have 35 computers running systems in our clean room.
Stuff like the Ellipsometer, PECVD,Etchers, SEM, three different RTA's, Furnaces.
How do you back up and protect all these systems from a disk failure?
We use WIN95, 98, and XP....whatever the manufacturers were using when they wrote the operating software.
cordially,
Keith Bradshaw
University of Texas at Dallas
_______________________________________________
labnetwork mailing list
labnetwork at mtl.mit.edu<mailto:labnetwork at mtl.mit.edu>
https://www-mtl.mit.edu/mailman/listinfo.cgi/labnetwork
<ATT00001..txt>
From whipp003 at umn.edu Fri Oct 15 23:03:21 2010
From: whipp003 at umn.edu (tony whipple)
Date: Fri, 15 Oct 2010 22:03:21 -0500
Subject: [labnetwork] System backups