> We did contact Bruker Nano and SFR (SFR is the company in Canada that
|
> offer service for Bruker Nano) and we?ve been said that Bruker does not
|
> anymore offer service or parts on old tools such as our Dektak 150 (tool
|
> has been bought in 2008, running on WinXP).
|
>
|
> By any chance, do you know a third party company that can offer services
|
> on Bruker Nano Profilometer?
|
>
|
>
|
>
|
> Thank you
|
>
|
>
|
>
|
> Best
|
>
|
> Chris
|
>
|
>
|
>
|
>
|
>
|
> *Christophe Cl?ment*
|
>
|
> *Technicien laboratoire*
|
>
|
> Laboratoire de microfabrication (LMF)
|
>
|
> Groupe des Couches Minces (GCM) *www.gcmlab.ca <http://www.gcmlab.ca/>*
|
>
|
> Ecole Polytechnique de Montr?al www.polymtl.ca
|
>
|
> D?partement de g?nie physique
|
>
|
> * 2900 Boulevard Edouard Monpetit
|
>
|
> Pavillon JAB
|
> Campus de l'Universit? de Montr?al
|
> Montr?al (Qu?bec) H3T 1J4
|
>
|
>
|
> 8 christophe.clement at polymtl.ca
|
>
|
> ( 514 340 4711 #2417
|
>
|
> Fax : 514 340 5195
|
>
|
>
|
>
|
>
|
> _______________________________________________
|
> labnetwork mailing list
|
> labnetwork at mtl.mit.edu
|
> https://mtl.mit.edu/mailman/listinfo.cgi/labnetwork
|
>
|
-------------- next part --------------
|
An HTML attachment was scrubbed...
|
URL: <https://mtl.mit.edu/pipermail/labnetwork/attachments/20231003/6e90846b/attachment.html>
|
From deonc69 at illinois.edu Tue Oct 3 16:04:54 2023
|
From: deonc69 at illinois.edu (Collins, Deon)
|
Date: Tue, 3 Oct 2023 20:04:54 +0000
|
Subject: [labnetwork] Deuterated Silane detection
|
In-Reply-To: <CWXP265MB48764A20668E33DF0B0CA71EB0C4A@CWXP265MB4876.GBRP265.PROD.OUTLOOK.COM>
|
References: <CWXP265MB48764A20668E33DF0B0CA71EB0C4A@CWXP265MB4876.GBRP265.PROD.OUTLOOK.COM>
|
Message-ID: <DM6PR11MB31624BBD0ED9744E1A4418E9D7C4A@DM6PR11MB3162.namprd11.prod.outlook.com>
|
DOD offers an electrochemical detection system. I believe your suspicions are correct and the SiD4 will has some level of cross detection and will probably be very close SiH4. We use PS-7's with assorted electrochemical cartridges in different areas for different materials. We have 86 of these units deployed within our facility. I have used the chart attached for detection purposes when I don't have the Propper cartridges available in my mobile detection units.
|
Deon D. Collins
|
FACILITY MANAGER
|
The Grainger College Of Engineering UIUC
|
Holonyak Micro & Nanotechnology Lab
|
208 N Wright St Rm. 1114 | MC-249
|
Urbana, IL 61801
|
217-300-7531 | deonc69 at illinois.edu<mailto:deonc69 at illinois.edu>
|
[cid:image001.png at 01D9F609.A6A3EE00]<http://illinois.edu/>
|
Under the Illinois Freedom of Information Act any written communication to or from university employees regarding university business is a public record and may be subject to public disclosure.
|
Life is not about watching other people live it. It's about you living your own!
|
From: labnetwork <labnetwork-bounces at mtl.mit.edu> On Behalf Of Owain Clark
|
Sent: Tuesday, October 3, 2023 7:50 AM
|
To: labnetwork at mtl.mit.edu
|
Subject: [labnetwork] Deuterated Silane detection
|
Dear all, our researchers wish to add a SiD4 (Deuterated silane) line to the facility. Looking at the gas properties it is unsurprisingly incredibly similar to SiH4 as one might expect. And therefore one might also suspect that a SiH4 gas detection head would have near 100% cross sensitivity to it. However our gas detection system provider (Drager, polytron heads) cannot confirm this and has no data on the subject. Our safety staff are unsurprisingly not keen to proceed on the basis of "it should work".
|
Does anyone run this gas in their facility and if so would you mind sharing information on your detection system for it? Or if anyone knows of a company that can fabricate a 4-20mA capable gas detection head pre-calibrated for it? Or as a 3rd line of inquiry if anyone knows of a service whereby one of our SiH4 standard polytron heads could be cross-calibrated against SiD4 empirically that would also be of interest.
|
BR, Owain
|
-------------- next part --------------
|
An HTML attachment was scrubbed...
|
URL: <https://mtl.mit.edu/pipermail/labnetwork/attachments/20231003/9536bed3/attachment.html>
|
-------------- next part --------------
|
A non-text attachment was scrubbed...
|
Name: image001.png
|
Type: image/png
|
Size: 2602 bytes
|
Desc: image001.png
|
URL: <https://mtl.mit.edu/pipermail/labnetwork/attachments/20231003/9536bed3/attachment.png>
|
-------------- next part --------------
|
A non-text attachment was scrubbed...
|
Name: PS7 Interferencegases Rev 12 12-04-17.pdf
|
Type: application/pdf
|
Subsets and Splits
No community queries yet
The top public SQL queries from the community will appear here once available.