text
stringlengths
1
2.97k
>> https://mtl.mit.edu/mailman/listinfo.cgi/labnetwork
>>
> --
> Best Regards,
> Dave Terry
> ???????????????????
> Sent from Gmail Mobile
> 617 784 7942
>
--
*Thank you, Jim*
*James Mitchell*
*Specialty Trades Technician*
*Dept. of Electrical and Computer Engineering*
*North Carolina State University Nanofabrication Facility (NNF)*
*MRC RM243A **Box 7911*
*2410 Campus Shore Dr., Raleigh, NC 27606*
*jtmitch5 at ncsu.edu* <jtmitch5 at ncsu.edu>
*Desk: 919-515-5394*
*Cell: 919-717-7325*
-------------- next part --------------
An HTML attachment was scrubbed...
URL: <https://mtl.mit.edu/pipermail/labnetwork/attachments/20230323/ff04fbd0/attachment.html>
From dave101260 at gmail.com Thu Mar 23 08:52:58 2023
From: dave101260 at gmail.com (Dave Terry)
Date: Thu, 23 Mar 2023 08:52:58 -0400
Subject: [labnetwork] Suss MA6 Question
In-Reply-To: <CAEXS_M-dtLLPA_M5TEijSeeKmiCBV3VhPhUcosXYbf+BYtjVAA@mail.gmail.com>
References: <CAEXS_M-9AfbQv1TzikGbZ_tLVdF4hf_O_rVA3Qh7wp2t4vXyKg@mail.gmail.com>
<CAF8gEbTGjqfL1HwPioX2v2eu_TCsPPwQNgNFfaF+f-A2O5a51g@mail.gmail.com>
<CAEXS_M-dtLLPA_M5TEijSeeKmiCBV3VhPhUcosXYbf+BYtjVAA@mail.gmail.com>
Message-ID: <CAF8gEbQFOLOXdh=LwKhywgO00kKCG_-9AyZpdRKCTwB4NoC8rw@mail.gmail.com>
It?ll take me a few to dig it up.
Give me a bit.
On Thu, Mar 23, 2023 at 8:50 AM James Mitchell <jtmitch5 at ncsu.edu> wrote:
> Dave, Do you have a part number for CB?
> I think I will try this first.
>
> Jim
>
> On Wed, Mar 22, 2023 at 7:17?PM Dave Terry <dave101260 at gmail.com> wrote:
>
>> I had a similar issue several years ago, and turned out to be the CB on
>> the back of the system.
>>
>> Swapped it out, and all is fine since.
>>
>>
>>
>> On Wed, Mar 22, 2023 at 3:48 PM James Mitchell <jtmitch5 at ncsu.edu> wrote:
>>
>>> Lamp power supply will not turn on after lamp failure.
>>> We had a lamp (350 W) pop on our system.
>>> Cleaned up and installed a new lamp.
>>> When we start the power supply the breaker in the back of the system
>>> trips.
>>> Specifically, it trips when we press CP. System starts at STAND-BY.
>>> Press "ON" and the display shows "READY". Press CP and it displays
>>> "Wait..." for a second and then trips.
>>> We checked for shorts and all looked good. We swapped to the other 220
>>> Volt plug in the back and it trips as well.
>>> Any ideas on this? We are not sure if the start box or the supply
>>> itself is the issue.
>>>
>>>
>>> *Thank you, Jim*
>>>
>>> *James Mitchell*
>>> *Specialty Trades Technician*
>>> *Dept. of Electrical and Computer Engineering*
>>> *North Carolina State University Nanofabrication Facility (NNF)*
>>> *MRC RM243A **Box 7911*
>>> *2410 Campus Shore Dr., Raleigh, NC 27606
>>> <https://www.google.com/maps/search/2410+Campus+Shore+Dr.,+Raleigh,+NC+27606?entry=gmail&source=g>*
>>> *jtmitch5 at ncsu.edu* <jtmitch5 at ncsu.edu>
>>> *Desk: 919-515-5394*
>>> *Cell: 919-717-7325*
>>>
>>>
>>> _______________________________________________
>>> labnetwork mailing list
>>> labnetwork at mtl.mit.edu
>>> https://mtl.mit.edu/mailman/listinfo.cgi/labnetwork
>>>
>> --
>> Best Regards,
>> Dave Terry
>> ???????????????????
>> Sent from Gmail Mobile
>> 617 784 7942
>>
>
>
> --
> *Thank you, Jim*
>
> *James Mitchell*
> *Specialty Trades Technician*
> *Dept. of Electrical and Computer Engineering*
> *North Carolina State University Nanofabrication Facility (NNF)*